Recognoil – User‘s Manual
Software version 2.2.1_2015-02-21
-
25
-
6. VALIDATION - CALIBRATION
(Professional version only)
In case of Recognoil
Professional version, an
aluminium sample validation-
calibration sheet has been
delivered to you, allowing
you to perform validation
test periodically.
Validation procedure can be
executed from the Settings
menu.
Please follow the
instructions of every
step.
In the final step, the
resulting value is shown. If
the value is much bigger
than the preset sample
value, it is necessary to clean
the device. Please click
“Cleaning instructions”.
If the value is smaller, it is
recommended to perform
calibration, clicking the
Calibration button.