PROVASET T3LPF REV.20210610
28
3.12.2 Flow rate test: connection with local pressure reading
In order to simplify the connection to the component being tested, you can also use
compressed air supply input
D>E
bypass for the local reading of the test pressure
J component being tested
CD
E
G
H
F
TEST
p
J
The pressure sensor (input port H) must be connected to the output port to read
the pressure inside the equipment.
The connection is simpler, however the test pressure is measured locally at the TEST
output port, inside the equipment, not directly on the component being tested.
In this case, the test pressure reading also includes the pressure drop , due to the
, through the connection tube, down to the input port
of the component being tested.