PROVASET T3LPF REV.20210610
76
The test cycle is performed with the parameters of the active test program.
already pro-
grammed in the test parameters.
VOLUMETRIC COEFFICIENT
Run a test WITHOUT the leak master:
START
TEST
PROG n°001: VALVE_A
On the next screen click on START TEST, being careful not to connect the calibrated
leak for now.
VOLUMETRIC COEFFICIENT
Repeat or go to next step
START
TEST
PROG n°001: VALVE_A
NEXT
STEP
measurement just performed and proceed with the next step or repeat from the
beginning.
VOLUMETRIC COEFFICIENT
PROG n°001: VALVE_A
Please insert the leak master
Master Leak scc/m
0.000
Choosing to continue, you will have to connect the leak master in any point of the
pneumatic circuit, in derivation on the pipe connected to the TEST output or on the