13. ALARMS
If there are alarms active, a new test cannot be started.
A test in progress is automatically terminated if an alarm occurs and the result of the
test is not already acquired, including marking, if envisaged.

Data memory alarm: the test parameters are not valid; check and reprogram all the
test parameters of the active product.


and reprogram all the parameters of the menu.


valid; repeat the calibration procedure.

Disconnect the compressed air supply, discharge the pressure in the test circuit.
 In case of out of scale pressure, follow a safety procedure to manual-
ly discharge the pressure into the test pneumatic circuit and into the component
being tested.

Flow sensor alarm: reading error.
If the error persists, technical assistance is required.

Pressure not discharged at the end of test.
 this alarm occurs when the machine has failed to release the pres-
sure of the piece under test having set the FSM parameter on PRESSURE. If the
FSM parameter is not set to pressure (for example: to TIME or WAITING) and
the machine for some reason has failed to unload, the alarm will not appear in
this case.
If a new test cannot be started due to an alarm, the alarm type is displayed on a red
background.
The active alarms are visible in a dedicated page of the base menu, which can be ac-
cessed by pressing the key
.