EasyManua.ls Logo

Tektronix 422 - SHORT-FORM PROCEDURE

Tektronix 422
213 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Check/CalibrationType 422 AC-DC
SHORT-FORM PROCEDURE
Type 422, Serial No. _______________
Calibration D ate
-------------------------------
Calibrated B y
------------------------------------
1. Adjust -12-Volt Power Supply (R1130) Page 5-9
REQUIREMENT: -1 2 volts ±0.12 volt.
PERFORMANCE: -1 2 volts ± _____vo lt
2. Adjust Blocking Oscillator Frequency Page 5-9
(R1125)
REQUIREMENT: Correct oscillator frequency; see
complete procedure.
PERFORMANCE: Correct
___
; incorrect
____
.
3. Check DC Regulation Page 5-10
Supply
REQUIREMENT:
PERFORMANCE:
+10.5 volt
+10.3 to +10.7 volts
volts
-1 2 volt
-11.9 to -12.1 volts
volts
+12 volt
+11.7 to +12.3 volts
vblts
+55 volt
+53.9 to +56.7 volts
volts
+95 volt +93.7 to +97.9 volts
volts
81 volt 80 to 82 volts
volts
4 Check DC Ripple
Page 5-11
Supply
REQUIREMENT:
(maximum)
PERFORMANCE:
-1 2 volt
20 millivolts
millivolts
+12 volt
20 millivolts
millivolts
+55 volt
0.32 volt
volt
+95 volt
0.4 volt
volt
+10.5 volt
1 millivolt
millivolt
81 volt
20 millivolts
millivolts
5. Adjust Low-Voltage Indicator (R1047) Page 5-12
REQUIREMENT: Power light begins to blink with DC
input voltage of 22 volts +0, 1 volt.
PERFORMANCE: Correct
___
; incorrect
____
6. Check AC Operation Page 5-12
Output voltage
Supply
REQUIREMENT:
PERFORMANCE:
-1 2 volt
11.9 to 12.1 volts
volts
+12 volt +11.7 to +12.3 volts
volts
+55 volt
+53.9 to +56.7 volts
volts
+95 volt
+93.7 to +97.9 volts
volts
+10.5 volt +10.3 to +10.7 volts volts
81 volt
80 to 82 volts
volts
Ripple
Supply
REQUIREMENT:
(maximum)
PERFORMANCE:
-1 2 volt
8 millivolts
millivolts
+12 volt
8 millivolts
millivolts
+10.5 volt 1 millivolt
millivolt
81 volt
8 millivolts
millivolts
7. Check Battery Charger Operation Page 5-13
REQUIREMENT:
CHARGE BATT 115V AC; 95-volt line, 325
milliamperes minimum. 126.5-volt line, 450 milliamperes
maximum.
OPERATE 115V AC; 95-and 126.5 volt line, 30 to 50
milliamperes.
PERFORMANCE:
CHARGE BATT 115V AC; 95-volt line,
____
milli-
amperes. 126.5-volt line,
-------
milliamperes.
OPERATE 115V AC; 95- and 126.5-volt line,
--------
and
--------
milliamperes.
8. Check High-Voltage Supply Page 5-13
REQUIREMENT: -1385 volts ±69.5 volts.
PERFORMANCE: -1385 v o lts ±
____
volts.
9. Adjust unblanking Center (R869) Page 5-13
REQUIREMENT: Maximum trace intensity.
PERFORMANCE: Correct
-------
; incorrect----------
10. Check/Adjust Trace Rotation (R851) Page 5-13
REQUIREMENT: Trace parallel to center horizontal iine
within 0.1 division/10 divisions.
PERFORMANCE: W ithin
____
division.
11. Check/Adjust Y Axis Alignment (R856) Page 5-14
REQUIREMENT: Markers parallel to the center vertical
line within 0.1 division.
PERFORMANCE: W ithin
------
division.
12. Check/Adjust CRT Geometry (R854) Page 5-15
REQUIREMENT: Bowing of markers at left and right
edges of graticule within 0.1 division. Trace parallel to the
top and bottom horizontal lines of the graticule within 0.1
division.
PERFORMANCE: Left and right edges within
division; top and bottom w ithin
____
division.
H i
5-3

Related product manuals