EasyManuals Logo
Home>Tektronix>Test Equipment>475

Tektronix 475 User Manual

Tektronix 475
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
Calibration475
c. CHECKThat the CH 1 UNCAL light turns on when
the VAR control is out of the detent position.
d. CHECKCRT display for 2 divisions or less of trace
shift when rotating the CH 1 VAR control through its
range.
e. ADJUSTCH 1 Variable Balance adjustment, R110
(see Fig. 5-8) for minimum trace shift when rotating the
CH 1 VAR control through its range.
f. Return the CH 1 VAR control to the detent position.
15. Adjust CH 1 Gain Switch Balance
a. Set the CH 1 VOLTS/DIV switch to 2 mV.
b. Position the trace to the center horizontal line.
c. CHECKCRT display for minimum trace shift
between 2 mV and 5 mV positions of the CH 1 VOLTS/
DIV switch.
d. ADJUSTCH 1 Gain Switch Bal adjustment, R135
(see Fig. 5-8), for minimum trace shift between 2 mV and
5 mV positions of the CH 1 VO LTS/DIV switch.
16. Check Probe Indicator Lights
a. Set both VOLTS/DIV switches to 5 mV.
b. Connect a X I0 probe to the CH 1 input.
c. CHECKLight under 5 mV setting is extinguished
and the light under 50 mV setting turns on.
Fig. 5-8. Location of CH 1 Var Voits/Div Balance and Gain Switch
Balance adjustments.
d. Set the VERT MODE switch to CH 2.
e. Move the X I0 probe to the CH 2 input.
f. CHECKLight under 5 mV setting is extinguished and
the light under 50 mV setting turns on.
g. Remove the X I0 probe.
17. Check AC-GND-DC Switches
a. Set both AC-GND-DC switches to DC.
b. Connect the Standard Amplitude Calibrator output to
the CH 2 input via a 42-inch 50 12 BNC cable.
c. Adjust the Standard Amplitude Calibrator for a
20 mV output.
d. Position the bottom of the display to the center
horizontal line.
e. Set CH 2 AC-GND-DC switch to GND.
f. CHECKFor no vertical deflection, trace is at the
center horizontal line.
g. Set the CH 2 AC-GND-DC switch to AC.
h. CHECKThat the display is centered about the
center horizontal line.
i. Move the test signal to the CH 1 input.
j. Set the VERT MODE switch to CH 1.
k. Position the bottom of the display to the center
horizontal line,
l. Set the CH 1 AC-GND-DC switch to GND.
m. Repeat step 17-f.
n. Set the CH 1 AC-GND-DC switch to AC.
®
5-30

Other manuals for Tektronix 475

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 475 and is the answer not in the manual?

Tektronix 475 Specifications

General IconGeneral
BrandTektronix
Model475
CategoryTest Equipment
LanguageEnglish

Related product manuals