Performance verification
Frequency, Amplitude, DC Offset, and AC Flatness Test Record (cont.)
Frequency 50.00 MHz
Reference - 0.50 dB
dB
Reference + 0.50 dB
Frequency 100.0 MHz
Reference - 1.00 dB
dB
Reference + 1.00 dB
Frequency 150.00 MHz
Reference - 1.00 dB
dB
Reference + 1.00 dB
Frequency 240.0 MHz
Reference - 2.00 dB
dB
Reference + 2.00 dB
Harmonic Distortion Test Record
Harmonic Distortio n
(amplitude 1 V
p-p
)
Fundamental
= reference 2nd 3rd 4th 5th Limit
AFG3011 / 3011C Spectrum A nalyzer reading
Sine 20 kHz
20 kHz 40 kHz 60 kHz 80 kHz 100 kHz
CH1 Harmonic
Disto
rtion (dBc)
reading -
refer
ence
(dBc)
0 dBc
Nth - r
eference <
-55 dBc
Sine 100 kHz
100 k
Hz
200 kHz 300 kHz
400 k
Hz
500 k
Hz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dB
c)
0 dBc
Nth-reference<
-55
dBc
Si
ne 1 MHz
1MHz 2MHz 3MHz 4MHz 5MHz
CH
1 H armonic
Distortion (dBc)
r
eading -
reference
(dBc)
0
dBc
Nth-reference<
-45 dBc
Sine 10 MHz
10 MHz 20 MHz 35 MHz 40 MHz 50 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-45 dBc
AFG302xB / 302xC Spectrum A nalyzer reading
Sine 20 kHz
20 kHz 40 kHz 60 kHz 80 kHz 100 kHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
AFG3000C and AFG3000 Series Specifications and Performance Verification 37