EasyManuals Logo

Tektronix DMM4040 User Manual

Tektronix DMM4040
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
DMM4040/4050
Users Manual
D-2
Table D-2. Ohms Analog Filter Errors
Range NPLC Additional Analog Filter Error
10, 100 0.5 mΩ
10 Ω
<10 1.9 mΩ
10, 100 1.5 mΩ
100 Ω
<10 9.0 mΩ
10, 100 0.6 Ω
100 kΩ
<10 2.5 Ω
Table D-3. DC Current Analog Filter Errors
Range NPLC Additional Analog Filter Error
100 μA, 10 mA, 1 A 100 0.005 % of range
10 0.015 % of range
1 0.027 % of range
0.2 0.09 % of range
0.02 0.27 % of range
1 mA, 100 mA, 10 A
[1]
10 0.001 % of range
1 0.0025 % of range
0.2 0.009 % of range
0.02 0.026 % of range
[1] 3 A range uses the 10 A range errors.