EasyManuals Logo
Home>Tektronix>Multimeter>DMM912

Tektronix DMM912 Technical Reference

Tektronix DMM912
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
DTM500 Series Specifications
Handheld and Benchtop Instruments Basic Service
3
Table 1: General Characteristics (Cont.)
Characteristic Description
nAccuracy
Temperature Range Resolution % of Reading
D
T
M
5
2
0
(
K-
t
y
pe probe) –200_ to100_ C 0.1_ C ±(0.3% + 1_ C)
–328_ to148_ F 0.2_ F ±(0.3% + 2_ F)
–99.9_ to 999.9_ C 0.1_ C ±(0.1% + 0.7_ C)
–147.9_ to 999.9_ F 0.2_ F ±(0.1% + 1.4_ F)
1000_ to 1370_ C 1_ C ±(0.3% + 1_ C)
1000_ to 2498_ F 2_ F ±(0.3% + 2_ F)
T1 – T2 reading
±(0.3% + 2.2_ C)
nAccuracy
Temperature Range Resolution % of Reading
DTM520 (J-type probe) –200_ to100_ C 0.1_ C ±(0.3% + 1.1_ C)
–328_ to148_ F 0.2_ F ±(0.3% + 2.2_ F)
–99.9_ to 760_ C 0.1_ C ±(0.1% + 0.8_ C)
–147.9_ to 999.9_ F 0.2_ F ±(0.1% + 1.6_ F)
1000_ to 1400_ F 2_ F ±(0.3% + 2_ F)
Table 2: General Characteristics
Characteristic Description
Temperature Range ATP01 bead probe: –40_ to 204_ C
Probe Tolerance ATP01 bead probe: ±2.2_ C
Measurement Rate (Readings/second)
DTM510 2.5
DTM520 (T2 or T1) 1
DTM520 (T1 – T2) 0.5
Input Protection 24 V maximum. Class III as defined in IEC 1010, Safety Requirements for Electrical
equipment for measurement, control, and laboratory use. Class III equipment is
equipment for connection to SELV or SELV-E circuits only.
Electrical Isolation T1 to T2 20 k minimum

Table of Contents

Other manuals for Tektronix DMM912

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DMM912 and is the answer not in the manual?

Tektronix DMM912 Specifications

General IconGeneral
BrandTektronix
ModelDMM912
CategoryMultimeter
LanguageEnglish

Related product manuals