EasyManuals Logo
Home>Tektronix>Test Equipment>DPO2012B

Tektronix DPO2012B Programmer's Manual

Tektronix DPO2012B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
Commands Listed in Alphabetical Order
:SAVE:WAVEFOR
M:FILEFORMAT SPREADSHEET; :SAVE:WAVEFORM ALL,
"E:/test_folder/test1_all. csv"
saves all displayed waveforms
(excluding serial bus waveforms) to
E:/test_folder/test1_all.cs v.
:SAVe:WAVEform:FILEFormat INTERNal; :SAVe:WAVEform ALL,
"E:/test_folder/test1"
saves all displayed waveforms (excluding serial bus
waveforms) to individual les named
E:/test_folder/test1_<wfm>.i sf
(for example test1_CH1.isf).
SAVe:WAVEform:FILEFormat
Species
or returns the le format for saved waveforms. Waveform header and
timing information is included in the resulting le of non-internal formats.
The oscilloscope saves DPO waveforms as a 500 x 200 matrix, with the rst
row corresponding to the most recently acquired data. The values specied by
DATa:STARt and DATa:STOP determinetherangeofwaveformdatatooutput.
In the event that DATa:STOP value is greater than the current record length, the
curren
t record length determines the last output value.
Group
Save and Recall
Syntax
SAVe:WAVEform:FILEFormat {INTERNal|SPR EADSheet}
SAVe:WAVEform:FILEFormat?
Related Commands
CURVe, DATa, DATa:STARt, DATa:STOP, SAVe:WAVEform,
WFMInpre:NR_Pt, WFMOutpre:NR_Pt?
Arguments
INTERNal speciesthatwaveformsaresavedinaninternalformat,usinga.isf
lename extension. These les can be recalled as reference waveforms. When this
argument is specied, the settings specied via the DATa:STARt and DATa:STOP
commands have no mea ning a s the entire waveform is saved.
SPREADSheet species that waveform data is saved in a format that contains
comma delimited values. These waveform data les are named using the .csv
lename extension. Saving waveforms in CSV format enables spreadsheet
programs to import the d ata.
Examples
SAVE:WAVEFORM:FILEFORMAT INTERNAL species that the inter nal le format
is the format used for saving waveforms.
SAVE:WAVEFORM:FILEFORMAT? might return :SAVE: WAVEFORM:FILEFORMAT
INTERNAL
indicating that waveforms are saved using the internal format.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-239

Other manuals for Tektronix DPO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012B and is the answer not in the manual?

Tektronix DPO2012B Specifications

General IconGeneral
BrandTektronix
ModelDPO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals