Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:PULSEWidth:SOUrc e {{CH1|CH2|CH3|CH4|
D0|D1|D2|D3|D4|D5|D6|D7|D8 |D9|D10|D11|D12|D13|D14|D15|LINE|EXT}
TRIGger:A:PULSEWidth:SOUrc e?
Arguments
CH1–CH4 specifies an analog input channel as the A edge trigger source.
D0–D15 specifies an digital input channel as the A edge trigger source.
EXT specifies an external trigger using the Aux In connector located on the front
panel of the oscilloscop
e.
LINE specifies AC line voltage.
Examples
TRIGGER:A:PULSEWIDTH:SOURC E CH1 sets channel 1 as the pulse width source.
TRIGGER:A:PULSEWIDTH:SOURC E? might return :TRIGGER:A:P ULSEWIDTH:
SOURCE CH1
indicating that channel 1 is the pulse width source.
TRIGger:A:PULSEWidth:WHEn
Sets or returns whether to trigger on a pulse that meets, falls outside, or within
the specified range of limits.
Group
Trigger
Syntax
TRIGger:A:PULSEWidth:WHEn {LESSthan|MO REthan|EQual|UNE Qual}
TRIGger:A:PULSEWidth:WHEn?
Related Commands
TRIGger:A:PULSEWidth:WIDth
Arguments
LESSthan argument sets the oscilloscope to trigger if a pulse is detected with
width less t
han the time set by the TRIGger:A:PULSEWidth:WIDth command.
MOREthan argument sets the oscilloscope to trigger if a pulse is detected with
width m ore than the time set by the TRIGger:A:PULSEWidth:WIDth command.
EQUal argument sets the oscilloscope to trigger if a pulse is detected with width
equal to the time period specified in TRIGger:A:PULSEWidth:WIDth within
a ±5% tolerance.
UNEQual argument sets the oscilloscope to trigger if a pulse is detected with
width greater than or less than (but not equal) the time period specifi ed in
TRIGger:A:PULSEWidth:WIDth within a ±5% tolerance.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-329