Commands Listed in Alphabetical Order
Examples
WFMINPRE:COMP
OSITION?
might return :WFMINPRE:COMPOSI TION
SINGULAR_YT
indicating that the sample part of singular waveform is transferred
to the instrument.
WFMINPRE:COMPOSITIONCOMPO SITE_YT sets the sample part of composite
waveform to be transferred to the instrument.
WFMInpre:ENCdg
Sets or returns the type of encoding for incoming waveform dat
a.
Group
Waveform Transfer
Syntax
WFMInpre:ENCdg {ASCii|BINary}
WFMInpre:ENCdg?
Related Commands
WFMOutpre:ENCdg
Arguments
ASCii specifies that the incoming data is in ASCII format.
BINary specifies that the incoming data is in a binary format whose further
interpretation requires knowledge of BYT_NR, BIT_NR, BN_FMT, and
BYT_OR.
Examples
WFMINPRE:ENCDG ASC sets the format o f incoming waveform data to ASCII
format.
WFMINPRE:ENCDG ? might return :WFMINPRE:ENCDG BIN indicating that the
incoming waveform data is in binary format.
WFMInpre:FILTERFreq
Specifies or returns the FilterVu low pass filter frequency, which was applied to
the waveform being sent to the instrument.
Group
Waveform Transfer
Syntax
WFMInpre:FILTERFreq {NR1}
WFMInpre:FILTERFreq?
Arguments
<NR1> is an signed integer.
2-356 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual