Commands Listed in Alphabetical Order
Arguments
Execute auto ze
ros the probe.
CH<x>:PRObe:COMMAND (No Query Form)
Sets the state of the p robe control specified with the first argument to the state
specified with the second argument. The commands and states are unique to the
attached probe type. Only certain VPI probes support this command. See the
probe documentation for how to set these string arguments.
Group
Vertical
Syntax
CH<x>:PRObe:COMMAND <QString>, <QString>
Arguments
<QString> are quoted strings specifying the probe command and value to set in
the probe attached to the specified channel.
Examples
CH1:PROBE:COMMAND “MODE”, “4–4V1MHz” sets a Tektronix VPI-DPG probe
to the 4-4V1MHz m ode.
CH1:PROBE:COMMAND “OUTPUT”, “ON” turns the output of a Tektronix
VPI-DPG probe on.
CH1:PROBE:COMMAND?“MODE” might return CH1:PROBE:COMMAND
“MODE”,“4-4V1MHZ”
.
CH<x>:PRObe:DEGAUss (No Query Form)
Starts a degauss auto-zero cycle on a TekVPI current probe attached to the input
channel specified by <x>, where x is the channel number.
Group
Vertical
Syntax
CH<x>:PRObe:DEGAUss EXECute
Arguments
EXECute initiates the degauss operation.
CH<x>:PRObe:DEGAUss:STATE? (Query Only)
Returns the state of the probe degauss for the channel specified by <x>, where is x
is the channel number.
2-104 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual