Commands Listed in Alphabetical Order
SAVe:WAVEfor
m:GATIng
Specifies whether save waveform operations should save the entire waveform
(NONe) or a specified portion of the waveform.
Group
Save and Recall
Syntax
SAVe:WAVEform:GATIng {NONe|CURSors| SCREEN}
SAVe:WAVEform:GATIng?
Arguments
CURSors turns on cursors and the gates are the waveform record points at the
cursor positions.
NONe saves the entire waveform.
SCREEN, if zoom is on, the gates are the start and end waveform record points of
the zoom (upper) graticule, otherwise the gates are the start and end waveform
record points of the main graticule.
Examples
SAVE:WAVEFORM:GATING CURSors specifies that, when the waveform gating
is set to cursors, save waveform operations should save the waveform points
between the cursors. If cursors are turned off, waveform gating automatically
reverts to NONE.
SAVe:WAVEform:SPREADSheet:RESOlution
Specifies whether to save the full (LRL) or reduced (thumb) resolution waveform
to a CSV file.
Group
Waveform Enhanced
Syntax
SAVe:WAVEform:SPREADSheet :RESOlution {FULL|REDUced }
SAVe:WAVEform:SPREADSheet :RESOlution?
Arguments
FULL specifies that full resolution waveform is saved to a CSV file.
REDUced specifies that reduced resolution wave form is saved to a CSV file.
Examples
SAVE:WAVEFORM:SPREADSHEET :RESOLUTION? might return
SAVE:WAVEFORM:SPREADSHEET :RESOLUTIONFULL indicating that full
resolution waveform is saved to a CSV file.
2-240 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual