EasyManua.ls Logo

Tektronix DTM500 Series - Page 24

Tektronix DTM500 Series
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DCM300 and DCM320 Performance Verification
Handheld and Benchtop Instruments Basic Service
9
Table 9: Resistance Checks
Resistance
Calibrator Output
Tolerance Low Limit Test Result High Limit
0
±
9
000 009
Buzzer must sound
12
0
±11 109 131
Buzzer must sound
1.000 k ±0.029 k 971 1.029 k
3.700 k ±0.083 k 3.617 k 3.783 k
39.00 k ±0.87 k 38.13 k 39.87 k