TAS 200 Series Adjustment Procedures
Handheld and Benchtop Instruments Basic Service
25
Following the warm-up period, preset the oscilloscope to the settings listed
below.
Table 20: Oscilloscope Initial Settings
Control Setting
INTENSITY Visible display
CURSOR ON/OFF Off
VERTICAL
CH 1 (CH 2) POSITION Midrange
MODE CH1 (CH2)
CH 1 (CH 2) VOLTS/DIV 10 mV
CH 1 (CH 2) VOLTS/DIV VARIABLE CAL
CH 1 (CH 2) GND In
HORIZONTAL
POSITION Midrange
X10 MAG Out
X-Y Out
SWP UNCAL Out
TRIGGER
LEVEL Midrange
MODE AUTO
COUPLING AC
SOURCE CH 1 (CH 2)
SLOPE Rising (push
switch out)
Initial Settings