Commands Listed in Alphabetical Order
Syntax
SYSTem:MODUle
?
Returns
<>
where <string> is a description of the error.
Examples
SYSTEM:MODULE? might return ?, which means ?.
SYSTem:POSetup
This command determines how the power supply initializes when its power switch
is turned on. This command configures the instrument to power up with default
settings, or power up with the settings that were in effect when the instrument
was turned off.
Group
System
Syntax
SYSTem:POSetup {RST|RCL0}
SYSTem:POSetup?
Arguments
RST: initializes the power supply to default settings after a power cycle.
RCL0: saves the most recent settings and restores these after a power cycle.
Returns
RST: default settings are applied after a power cycle.
RCL0: most recent settings are saved and restored after a power cycle.
Examples
SYST:POS RST
SYSTEM:POSETUP?
might respond with RST, which would indicate that the
power supply is configured to restore the power supply to default settings when
it powers up.
SYSTem:REMote (No Query Form)
This command sets the power supply to remote control mode.
Group
System
Syntax
SYSTem:REMote
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-47