6½-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 2:
DMM6500-905-01 Rev. E June 2021 2-9
Verify the dc voltage 100 V range
The information in this section is intended for qualified service personnel only, as described
by the types of product users in the Safety precautions pages, provided at the beginning of
this document. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, that if contacted, could
cause personal injury or death. Use appropriate safety precautions when working with
hazardous voltages.
Description Nominal value Lower limit Upper limit
Verify the dc voltage 1000 V range
Description Nominal value Lower limit* Upper limit*
*For each additional volt over ±500 V, add 0.02 mV of uncertainty.