Section
2: Performance verification DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
2-36 DMM6500-905-01 Rev. E June 2021
Verify digitize current 100 µA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Full scale (+)
99.998
99.878
100.118
Half scale (+)
49.9991
49.9141
50.0841
Half scale (–)
-49.993
-50.0843
-49.9143
Full scale (–)
-99.9985
-100.1185
-99.8785
Verify digitize current 1 mA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Full scale (+)
0.999993
0.998993
1.000993
Half scale (+)
0.499998
0.499348
0.500648
Half scale (–)
-0.500009
-0.500659
-0.499359
Full scale (–)
-1.000017
-1.001017
-0.999017
Verify digitize current 10 mA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Full scale (+)
10.00004
9.99204
10.00804
Half scale (+)
5.00004
4.99454
5.00554
Half scale (–)
-5.00003
-5.00553
-4.99453
Full scale (–)
-10.00006
-10.00806
-9.99206
Verify digitize current 100 mA range
Description Calibrator setpoint
(A)
Lower limit Upper limit
Full scale (+)
100
99.92
100.08
Half scale (+)
50
49.945
50.055
Half scale (–)
-50
-50.055
-49.945
Full scale (–)
-100
-100.08
-99.92

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY DMM6500 and is the answer not in the manual?

Tektronix KEITHLEY DMM6500 Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY DMM6500
CategoryMultimeter
LanguageEnglish

Related product manuals