Section
2: Performance verification DMM6500 -Digit Multimeter with Scanning
Calibration and Adjustment Manual
2-48 DMM6500-905-01 Rev. E June 2021
Apply a relative offset to the DMM6500:
1. On the calibrator, select the OPR/STBY key. Ensure that the front panel displays STANDBY.
2. Verify the DMM6500 zero reading.
3. Verify the dc current verification points and reference DMM readings from the following table.
Verify dc current 10 A range
Description Nominal input Typical measured
input (A)
Lower limit Upper limit
Full scale (+) 10 9.99961 9.9755 10.0245
Half scale (+)
5
4.99984
4.9865
5.0135
Half scale (–)
–5
5.00079
-5.0135
-4.9865
Full scale (–)
10
10.0016
-10.0245
-9.9755