6½-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
DMM6500-905-01 Rev. E June 2021 3-9
Command parameters for a front-terminal adjustment with a 4-wire
short
When calibrating your DMM6500 for a front-terminal adjustment with a 4-wire short, use the following
command parameters.
Send each command parameter twice. First, send the parameter using the setup command. Second,
send the parameter using the execute command.
The example code below uses the first command parameter from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_DCV_100mV_zero_front")
cal.adjust.step.execute("cal_DCV_100mV_zero_front")
Command parameters
cal_3W_1kohm_Hi_zero_front
cal_3W_1kohm_SLO_zero_front
cal_3W_10kohm_SLO_zero_front
cal_2W_100kohm_zero_front