6½-Digit Multimeter with Scanning Calibration and Adjustment Manual Section 3:
DMM6500-905-01 Rev. E June 2021 3-23
Command parameters Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
10 µA range (+) full scale nominal
dc current 0.00022 0.00001 8508A or 8588A
10 µA
10 µA range (-) full scale nominal
dc current 0.00022 -0.00001
10 µA
100 µA range (+) full scale nominal
s
Reading
100 µA range (-) full scale nominal
cal_DCI_DIGI_100uA_fs_ne
g
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus value*
Reference
DMM dc
current range
1 mA range (+) full scale nominal
1 mA range (-) full scale nominal
10 mA range (+) full scale nominal