Commands Listed in Alphabetical Order
*OPC? might ret
urn 1 to indicate that all pending OPC operations are finished.
PASSWord(No Query Form)
Enables the *PUD and NEWpass set commands. Sending PASSWord without any
arguments disables these same commands. Once the password is successfully
entered, the *PUD and NEWpass commands are enabled until the oscilloscope
is powered off, or until the FACtory command or the PASSWord command with
no arguments is issued.
To change the password, you must first enter the valid password with the
PASSWord command and then c hange to your new password with the NEWpass
command. Remember that the password is case sensitive.
Group
Miscellaneous
Syntax
PASSWord <QString>
Related Commands
NEWpass, *PUD
Arguments
<QString> is the password and can incl
ude up to 10 characte rs. The factory
default password is “XYZZY” and is always valid.
Examples
PASSWORD “XYZZY” enables the *PUD and NEWpass set commands.
PASSWORD disables the * PUD and NEWpass set commands. You can still use the
query version of *PUD.
PICTBridge:DATEPrint
Enables or disables printing the date on the print output.
Group
PictBridge
Syntax
PICTBridge:DATEPrint {DEFLT|OFF|ON}
Arguments
DEFLT is the default setting of the printer.
ON enables the date print on the print output.
OFF disables the date print on the print output.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-221