Commands Listed in Alphabetical Order
Related Commands
DATE, TIME
Arguments
DATEOnly enables the display of date.
TIMEOnly enables the display of time.
DATETIME or ON enables the display o f both date and time.
NONe or OFF disables the display of date and time.
<NR1> = 0 disables the display of date and time; any other value enables the
display of date and time.
Examples
DISPLAY:CLOCK ON enables display of date and time.
DISPLAY:CLOCK? might return :DISPLAY:CLOCK 1 indicating that the display
of date and time is currently enabled.
DISplay:DIGital:HEIght
Sets or returns the number of available digital waveform position slots.
NOTE. If the number of live digital c hannels exceeds the number of slots, their
height is reduced to a smaller size, allowing all digital channels to be displayed.
Group
Display
Syntax
DISplay:DIGital:HEIght {SMAll|MEDium|L ARge}
DISplay:DIGital:HEIght?
Arguments
SMAll sets the height to 40.
MEDium sets the height to 20.
LARge sets the height to 10.
DISplay:FORMat
Sets or returns the display format.
Group
Display
2-140 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual