Commands Listed in Alphabetical Order
Examples
DIAG:SELECT:C
PU
sets the oscilloscope to run just CPU tests.
DIAg:STATE (No Query Form)
This command starts or stops the oscilloscope self-test. Depending on the
argument, self-test capabilities are either turned on or off.
Group
Calibration and Diagnostic
Syntax
DIAg:STATE {EXECute|ABORt}
Arguments
EXECute starts diagnostics.
ABORt stops diagnostics at the end of the current loop.
Examples
DIAG:STATE EXECute starts diagnostics.
DISplay? (Query Only)
Returns the current display settings.
Group
Display
Syntax
DISplay?
DISplay:CLOCk
Sets or returns whether the oscilloscope displays the date and time. The query
form of this command returns an ON (1) or an OFF (0).
NOTE. For the MSO/DPO2000/B series, the enumerations enable control of the
date/time display. The query form will return 0 when the control is set to NO
Ne or
OFF and will return 1 when the control is set to DATETIME or ON.
Group
Display
Syntax
DISplay:CLOCk {DATEOnly|TIMEOnly|DATET IME|NONe|ON|OFF| <NR1>}
DISplay:CLOCk?
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-139

Other manuals for Tektronix MSO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2024B and is the answer not in the manual?

Tektronix MSO2024B Specifications

General IconGeneral
BrandTektronix
ModelMSO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals