Command Descriptions
2–202
TDS Family Oscilloscope Programmer Manual
NEWpass, *PUD

 



 is the password and can include up to 10 characters. The factory
default password is “XYZZY”and is always valid.
 
Enables the *PUB and NEWpass set commands.

Disables the *PUD and NEWpass set commands. You can still use the query
version of *PUD.
*PSC
Sets and queries the power-on status flag that controls the automatic power-on
handling of the DESER, SRER, and ESER registers. When *PSC is true, the
DESER register is set to 255 and the SRER and ESER registers are set to 0 at
power-on. When *PSC is false, the current values in the DESER, SRER, and
ESER registers are preserved in nonvolatile memory when power is shut off and
are restored at power-on. For a complete discussion of the use of these registers,
see page 3–1.
Status and Error
DESE, *ESE, FACtory, *RST, *SRE
 




Related Commands
Syntax
Arguments
Examples
Group
Related Commands
Syntax