EasyManuals Logo

Tektronix TDS 500B Programming Manual

Tektronix TDS 500B
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #225 background imageLoading...
Page #225 background image
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–207
The *RST command does not alter the following:
H The state of the IEEE Std 488.1–1987 interface.
H The selected IEEE Std 488.1–1987 address of the digitizing oscilloscope.
H Calibration data that affect device specifications.
H The Output Queue.
H The Service Request Enable Register setting.
H The Standard Event Status Enable Register setting.
H The Power-on status clear flag setting.
H Alias definitions.
H Stored settings.
H The *PUD? response.
RS232:BAUd
RS-232/Centronics Hardcopy Interface Only
Sets or queries RS-232-C interface transmission speed.
RS232
 





 where  can be 300, 600, 1200, 2400, 4800, 9600 or 19200.
 
sets the transmission rate to 9600 baud.
Group
Syntax
Arguments
Examples