Command Descriptions
2–232
TDS Family Oscilloscope Programmer Manual
!
!
#"
enables the user to set the holdoff time.
#" automatically calculates a holdoff time to use. This time is typically
equivalent to the greater of ½ screen (5 divisions) of time or 250 ns. The
maximum value is 12 seconds. For example, if the oscilloscope is set to
1 msec/division then the default holdoff will be 1 msec/division × 25 divs =
25 msec.
set the holdoff to the by time setting. This enables the user to set the holdoff
time.
TRIGger:MAIn:HOLdoff:TIMe
TDS 684A & 7XXA Only
Sets or queries the main trigger holdoff time. This is equivalent to setting
Holdoff Time in the Mode & Holdoff side menu.
Trigger
!
!
!
holdoff time in seconds. The range is 250 ns to 12.0 seconds.
Arguments
Examples
Group
Syntax
Arguments