EasyManuals Logo

Tektronix TDS 500B Programming Manual

Tektronix TDS 500B
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #250 background imageLoading...
Page #250 background image
Command Descriptions
2–232
TDS Family Oscilloscope Programmer Manual
!

!

 

#"

 enables the user to set the holdoff time.
#" automatically calculates a holdoff time to use. This time is typically
equivalent to the greater of ½ screen (5 divisions) of time or 250 ns. The
maximum value is 12 seconds. For example, if the oscilloscope is set to
1 msec/division then the default holdoff will be 1 msec/division × 25 divs =
25 msec.
 
set the holdoff to the by time setting. This enables the user to set the holdoff
time.
TRIGger:MAIn:HOLdoff:TIMe
TDS 684A & 7XXA Only
Sets or queries the main trigger holdoff time. This is equivalent to setting
Holdoff Time in the Mode & Holdoff side menu.
Trigger
! 
!

!



 
 holdoff time in seconds. The range is 250 ns to 12.0 seconds.
Arguments
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS 500B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 500B and is the answer not in the manual?

Tektronix TDS 500B Specifications

General IconGeneral
BrandTektronix
ModelTDS 500B
CategoryTest Equipment
LanguageEnglish

Related product manuals