Performance Tests
1–74
TDS Family Performance Verification and Specifications
Digitizing Oscilloscope
50 Cable
Signal Generator
Output
Figure 1–31: 60 Hz Rejection Test Hookup
H Adjust the signal generator for three vertical divisions of 60 Hz
signal. See Figure 1–32. The signal will not be triggered. That is, it
will run free.
Figure 1–32: 60 Hz Rejection Test Setup Signal
c. Check 60 Hz rejection.
H Use the keypad to set the horizontal scale (/div) to 50 s (press 50,
SHIFT, m, then ENTER).
H Reconnect the output of the signal generator. Connect the composite
signal connector of the PAL signal source (labeled COMPST on the