Measuring Waveforms
3–104
TDS 500B, TDS 600B, & TDS 700A User Manual
2. Press Function (main) H Bars, V Bars, Paired, or Off (side).
Non-selected Cursor
(Dashed Vertical Bar)
Cursor Readout (Paired)
Selected Cursor
(Solid Vertical Bar)
Position of Vertical Bar Cursors (Useful for
Locating Cursors Outside the Display)
Figure 3–61: Paired Cursor Menu and Readouts
To select the cursor mode and adjust the cursors in either mode, do the following
steps:
1. Press CURSOR Mode (main) Independent or Track (side):
Independent makes each cursor positionable without regard to the position of
the other cursor.
Track makes both cursors positionable in tandem; that is, both cursors move in
unison and maintain a fixed horizontal or vertical distance between each other.
2. Adjust the cursors according to the mode you have selected:
To adjust either cursor in independent mode, use the general purpose knob to
move the selected (active) cursor. A solid line indicates the adjustable cursor
and a dashed line the fixed cursor. Press Select to toggle selection between
the two cursors.
To adjust both cursors in tracking mode, use the general purpose knob to
move both cursors.
Set Mode and Adjust
the Cursors