Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–293




 specifies that the waveform is transmitted as maximum and minimum point
pairs. Only y values are explicitly transmitted. Absolute coordinates are given
by:
X
n
0 XINcr (n PT_Off)
Y
n
max
YZEro YMUlt (y
n
max
YOFf)
Y
n
min
YZEro YMUlt (y
n
min
YOFf)
specifies a normal waveform where one ASCII or binary data point is
transmitted for each point in the waveform record. Only y values are explicitly
transmitted. Absolute coordinates are given by:
X
n
0 XINcr (n PT_Off)
Y
n
YZEro YMUlt (y
n
YOFf)
 
sets the waveform data point format to enveloped.
WFMPre:PT_Off (No Query Form)
Specifies the trigger point within the waveform record for the reference
waveform specified by the DATa:DESTination command.
Waveform
HORizontal:TRIGger:POsition
 


 
 = 0 to the recordlength, and is the position of the trigger point relative to
DATa:STARt.

specifies that the trigger point is the first point in the waveform record.
Arguments
Examples
Group
Related Commands
Syntax
Arguments
Examples