Performance Tests
1–66
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
Figure 1–25: Measurement of Probe Compensator Amplitude
g. Check against limits:
H Subtract the value just obtained (base level) from that obtained
previously (top level).
H CHECK that the difference obtained is within 495 mV to 505 mV,
inclusive.
H Enter voltage difference on test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.