Performance Tests
144
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
b. Match the trigger source to the channel selected:
H Press TRIGGER MENU. Press the main-menu button Source; then
press the side-menu button that corresponds to the channel selected.
c. Set its input impedance: (This substep is not required on the TDS694C
and TDS794D.)
H Press VERTICAL MENU. Press the main-menu button Coupling,
then press the side-menu W button to toggle it to the 50 W setting.
d. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 14 not yet checked. (Start with the 100 mV setting).
e. Set the triggering coupling:
H Press TRIGGER MENU.
H Press the main-menu button Coupling.
H For vertical settings of 1, 2, and 5 mV, press the side menu button
Noise Rej. For all other settings, press side menu button DC.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 700D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700D and is the answer not in the manual?

Tektronix TDS 700D Specifications

General IconGeneral
BrandTektronix
ModelTDS 700D
CategoryTest Equipment
LanguageEnglish

Related product manuals