Acquiring and Displaying Waveforms
3–72
TDS 500D, TDS 600B, & TDS 700D User Manual
Because FastFrame introduces additional processing time into the operation
cycle of acquire, process, and display, its best to use Single Sequence
Acquisition (see Acquire menu, Stop After menu). With Single Sequence
selected, you will see the current acquisition sequence; otherwise, the display
lags the current sequence by one sequence. You can also see the current
sequence by pressing the RUN/STOP button to stop the acquisition.
TimeStamp uses the measurement readouts; therefore, turning on FastFrame
TimeStamp turns off Measurements and vice versa.