2−163
When the DATa:SOUrce is not displayed, the TDS210 and TDS220
(firmware below V 2.00) with a TDS2CMA communications module
will return a value. All other oscilloscope, firmware version, and
module combinations will generate an error and will return event
code 2244.
Waveform
WFMPre:NR_Pt?
WFMPre:NR_Pt?
returns <NR1>, which is the number of data points. If DATa:WIDth is
2, then there are <NR1>*2 bytes in the curve.
DATa:ENCdg, DATa:SOUrce, DATa:STARt, DATa:STOP
WFMPre:PT_Fmt
The set form of this command sets the format (Y or ENV) of the
reference waveform specified by the DATa:DESTination command.
The query form returns the format of the waveform specified by the
DATa:SOUrce command, if that waveform is on or displayed. If the
waveform is not displayed, the query form of this command
generates an error and returns event code 2244.
Waveform
WFMPre:PT_Fmt { ENV | Y }
WFMPre:PT_Fmt?
Group
Syntax
Examples
Related Commands
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS1012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1012 and is the answer not in the manual?

Tektronix TDS1012 Specifications

General IconGeneral
BrandTektronix
ModelTDS1012
CategoryTest Equipment
LanguageEnglish

Related product manuals