Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−165
WFMPre:PT_Off
The set form of this command is ignored.
The query form always returns a 0, unless the DATA:SOUrce
waveform is not displayed, in which case the query generates an
error and returns event code 2244.
This command is included for compatibility with other Tektronix
oscilloscopes.
Waveform
WFMPre:PT_Off?
Arguments are ignored.
WFMPre:WFId? (Query Only)
Returns a descriptive string from the waveform specified in the
DATa:SOUrce command, if that waveform is active or displayed. If
that waveform is not active or displayed, the query fails and the
oscilloscope generates an execution error with event code 2244
(waveform requested is not active).
Waveform
WFMPre:WFId?
Group
Syntax
Arguments