-4-
C:C-Weighting. For checking the low-frequency
content of noise.
(If the C-Weighted level is much higher than the
A-weighted level, then there is a large amount of
low-frequency noise)
Time weighting select switch.
FAST:for normal measurements.
SLOW:for checking average level of fluctuation
noise.
Microphone
1/2 inch Electret Condenser microphone.
CAL potentiometer
Calibration control, For level calibration adjustment.
AC output terminal
1Vrms corresponding to each range step.
Output impedance≒50Ω
Output signal by standard 3.5mm coaxial socket signal
on pin.
DC output terminal
Output:10mV/dB
Output impedance≒100Ω
Output signal by standard 3.5mm coaxial socket signal
on pin.
External DC 9V power supply terminal
For connection with AC adapter.
Tripod mounting screw.
Battery cover
Calibration procedures
Using a standard Acoustic Calibrator(94dB,1KHz Sine
wave)
(1) Make the following switch settings.
Display:dBA
Time weighting:FAST
Measurement mode:MAX MIN Mode function
disable. Level range:50 to100dB