-4-
CC-Weighting. For checking the low-frequency
content of noise.
(If the C-Weighted level is much higher than the
A-weighted level, then there is a large amount of
low-frequency noise)
Time weighting select switch.
FASTfor normal measurements.
SLOWfor checking average level of fluctuation
noise.
Microphone
1/2 inch Electret Condenser microphone.
CAL potentiometer
Calibration control, For level calibration adjustment.
AC output terminal
1Vrms corresponding to each range step.
Output impedance50Ω
Output signal by standard 3.5mm coaxial socket signal
on pin.
DC output terminal
Output10mV/dB
Output impedance100Ω
Output signal by standard 3.5mm coaxial socket signal
on pin.
External DC 9V power supply terminal
For connection with AC adapter.
Tripod mounting screw.
Battery cover
Calibration procedures
Using a standard Acoustic Calibrator(94dB,1KHz Sine
wave)
(1) Make the following switch settings.
DisplaydBA
Time weightingFAST
Measurement modeMAX MIN Mode function
disable. Level range50 to100dB