STX-1 Series Instrument www.tensitron.com pg. 14
When calibrating these instruments tension values must be entered in
Lbs. Note: If your dead weight standards are in Kilograms, convert their
values into Lbs: 1 KG = 2.2046 Lbs
Follow these instructions to update the calibration for an existing strap on
the menu or to add a calibration for a new strap. If you would like to add a
new strap, first create a name for it by following the instructions under
“Custom Names” in Section 6, above.
1. Scroll to SETUP and press ENTER. Next, scroll to RE-CAL TENSION and press
ENTER. The screen will now display: SETUP, RE-CALIBRATE TENSION,
ENTER PASSWORD, CAL DATA WILL BE LOST!
2. Enter Password. Contact Tensitron at service@tensitron.com for the password.
3. The display now indicates: SETUP RE-CALIBRATE TENSION with a list of the
calibration names and custom names programmed into the instrument. Scroll to
the description you wish to recalibrate, highlight it and then press ENTER.