6/14
3 Package content
RaDEC indicator
Power adapter 230 VAC/5 VDC
USB power cable
Operation manual
4 Technical parameters
Product Air quality indicator RaDEC
Type designation 042 127 252 000 RaDEC
042 127 257 000 RaDEC light
Sensitivity for Rn concentration measurement 0.055 imp/hr/Bq/m
3
,
measured from RaA+RaC
Statistical error of Rn value at 300 Bq/m3 ± 24 % at 1 hour measurement
± 5 % at 24 h measurement
Rn measuring range approx. 30 – 60000 Bq/m
3
Rn measurement and display interval 6 minutes, (moving average 60x6 min)
Gamma detection sensitivity 0.75 imp/sec/1uSv/hr
Gamma measuring range 0.1 – 10000 uSv/hr
Gamma measurement and display interval 3 sec, value smoothed
By weighted averages
CO measuring range 0 –800 ppm
CO measurement and display interval 3 sec
CO measurement accuracy better than ± 20 % / 5 years
O
2
measuring range 0–40000 ppm
CO
2
measurement and display interval 1 min,
CO
2
measurement accuracy, ± 20 %
VOC measuring range 1–100 %
VOC measurement and display interval 3 sec
Temp. measuring range 40–125 °C
RH measuring range 0–100 %
Temp. and RH measurement and display interval 1 min
RH measurement accuracy ± 5 %
Max. number of records 2570
Recording interval 1 hour
Record value
radon current moving average value
other variables hourly average
Power supply built-in battery or
charger via USB-C port
Battery life min. 7 days
Dimension Ø 130 x 110 mm
Guaranteed operating conditions temperature: 10 to 50 °C
RH: 0 % to 80 %