EasyManua.ls Logo

TESTO 205 - Page 11

TESTO 205
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- If a stable reading is available (change less than 0.02pH in
20s), the calibration point calibrates and the instrument
changes to the next calibration point (if available) or to the
gradient and offset value display.
Carry out calibration manually:
CAL
.
Repeat Step 2 for additional calibration points.
- Once calibration is complete, the amount of the gradient and
offset value is shown. If the amount of the gradient value is less
than 50mV / pH or the amount of the offset value is greater
than 60mV, the pH electrode is spent and must be replaced.
3 Return to measurement view:
CAL
.
5. Operation 25
en
1.888.475.5235info@Testo-Direct.com
www.Testo-Direct.com

Other manuals for TESTO 205

Related product manuals