Using the product16
Using the product
²
Switching oon tthe iinstrument:
1 Press .
- All display segments light up (duration: 3 sec), the suction pump starts
(ventilator noise).
- The initialization phase is carried out (heating, auto-test). The remaining
duration is displayed.
- After the end of the initialization phase:
OK lights up (duration: 2 sec).
Attention! Danger of burning from the hot sensor head after prolonged use!
i Before touching the sensor head or packing up the instrument, switch off
and allow to cool.
²
Selecting ggas tto bbe ddetected ((only wwith ssensor hhead 00554 3180):
i Press several times, until the arrow lights up under the desired gas type.
For other gas types, please consult the table in the chapter Tips and assistance .
²
Carrying oout zzeroing:
i Keep pressed until two beep sounds are heard.
- The instrument is standardised to the current gas concentration.
²
Change ooperating mmode:
After switching on, the instrument is automatically in Localization mode (optimization
of the sensitivity to the change in the signal). If r
equired, Search mode can be
activated (optimization of the sensitivity to the leakage quantity).
i Press .
- disappears. Search mode is activated.
i Press again.
- lights up in display. Localization mode is activated again.
²
Carrying oout ggas ddetection:
Attention! Destruction of the sensor by non-desorbant substances (e.g. oils)!
i Do not use the instrument in dirty surroundings.
i Move the sensor head as close as possible, and slowly (3 to 5 cm per second)
over the parts which are to be examined for leaks.