6. Measuring 31
²
Resetting tthe mmaximum/minimum vvalues:
The maximum/minimum values of all channels can be
reset to the current reading.
1 Press several times until Max or Min lights up.
2 Press and hold (approx. 2s).
- All maximum or minimum values are reset to the
current reading.
²
Printing rreadings:
The readings shown on the display (current reading,
recorded reading or max./min. reading) can be printed
out.
A Testo printer is required (accessory part).
With the Max./Min. print function switched on, the
maximum and minimum values are printed out as well
as the current reading or recorded reading.
º See the chapter P
ERFORMING SETTINGS.
1 Configure the instrument so that the value to be
printed is shown on the display.
2 Press .