EasyManua.ls Logo

Textron Greenlee DM-200A - Frequency Range Accuracy Table; Frequency-Logic Level Accuracy Table; Non-contact EF Accuracy Table

Textron Greenlee DM-200A
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DM-200A • DM-210A • DM-510A
19
Accuracy (cont’d)
All Models (cont’d)
Accuracy of Frequency Ranges
Display Range Accuracy
10.00 Hz to 65.53 Hz ± (0.03% + 0.03 Hz)
65.5 Hz to 655.3 Hz ± (0.03% + 0.3 Hz)
0.655 kHz to 6.553 kHz ± (0.03% + 0.003 kHz)
6.55 kHz to 50.00 kHz ± (0.03% + 0.03 kHz)
Frequency—Logic Level Hz (mV Function)
Range Accuracy
Sensitivity
(square wave)
5.0 Hz to 6.553 Hz ± (0.03% + 0.002 Hz)
3 V peak
6.55 Hz to 65.53 Hz ± (0.03% + 0.02 Hz)
65.5 Hz to 655.3 Hz ± (0.03% + 0.2 Hz)
0.655 kHz to 6.553 kHz ± (0.03% + 0.002 kHz)
6.55 kHz to 65.53 kHz ± (0.03% + 0.02 kHz)
65.5 kHz to 500.0 kHz ± (0.03% + 0.2 kHz)
500.0 kHz to 655.3 kHz ± (0.03% + 0.2 kHz)
5 V peak
0.655 MHz to 1.000 MHz ± (0.03% + 0.002 MHz)
Non-contact Electric Field Detection (EF)
Typical Voltage Bar Graph Indication* Frequency Range
10 V to 36 V
50 Hz to 60 Hz
23 V to 83 V – –
59 V to 165 V – – –
124 V to 330 V – – – –
250 V to 1000 V – – – – –
*
Bar graph indication and tone are proportional to signal strength