Table 2
declaration - electromagnetic immunity
Immunity test IEC 60601 test level Compliance level
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Not applicable
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV line(s)
to lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
Not applicable
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0 % UT; 0.5 cycle at
0°, 45°, 90°, 135°, 180°,
225°, 270°and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300
cycles
Not applicable
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.