EasyManua.ls Logo

THERAGUN PRO PLUS - Page 21

THERAGUN PRO PLUS
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Document Name
User Manual
Document No. TG-TPP-001
Model
Theragun PRO PLUS
Rev
01
Page 21 of 24
Table 2
declaration - electromagnetic immunity
Immunity test IEC 60601 test level Compliance level
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Not applicable
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV line(s)
to lines
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
Not applicable
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0 % UT; 0.5 cycle at
, 45°, 90°, 13, 180°,
225°, 270°and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at
0 % UT; 250/300
cycles
Not applicable
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
NOTE: UT is the a.c. mains voltage prior to application of the test level.