Fluid Bags may be replaced as needed.
No button press required.
A. Close clamp on empty bag, open
clamp on full bag & remove empty
bag from hook
B. Remove spike from bag
C. Spike new bag
D. Hang new bag and open clamp
when fluid is needed
Note: For best accuracy minimize bouncing
on hooks.
Suspend fluid collection by disconnecting or
clamping suction source.
Failure to do so will
result in error in the displayed deficit
volumes.
Note: Fluid inflow may continue.
Select Change suction cans.
Remove, add, or replace canisters, tissue
collection accessories, or additional suction
lines as needed. Re-connect vacuum and/or
suction lines.
Do Not resume suction yet. Allow movement
or vibration of canister ring caused by
changing canisters and/or reconnecting
vacuum and/or suction lines to dissipate.