22 | P a g e
Part Number - 00824 Revision K, DCO 2017-0074
© 2017 Thermedx
®
, LLC
Bag Changes
Fluid Bags may be replaced as needed.
No button press required.
A. Close clamp on empty bag, open
clamp on full bag & remove empty
bag from hook
B. Remove spike from bag
C. Spike new bag
D. Hang new bag and open clamp
when fluid is needed
Note: For best accuracy minimize bouncing
on hooks.
Canister Changes
Suspend fluid collection by disconnecting or
clamping suction source.
Failure to do so will
result in error in the displayed deficit
volumes.
Note: Fluid inflow may continue.
Select Change suction cans.
Remove, add, or replace canisters, tissue
collection accessories, or additional suction
lines as needed. Re-connect vacuum and/or
suction lines.
Do Not resume suction yet. Allow movement
or vibration of canister ring caused by
changing canisters and/or reconnecting
vacuum and/or suction lines to dissipate.