Cat. No. 3593, 3594: The fuse drawer will have two
active fuses installed at the factory. Fuse is rated for
6.3 Amps (part no. 50607A).
Cat No. 3595: The fuse drawer will have one spare and
one active fuse installed at the factory. Fuse is rated for
8.0 Amps (part no. 50021).
Cat No. 3596: The fuse drawer will have one spare and
one active fuse installed at the factory. Fuse is rated for
10.0 Amps (part no. 49998). Note that the frequency
selector adjacent to the power entry module must be
set for the correct frequency.
Ensure that the fuses are securely in place and reinstall
the entire drawer into the power entry module.
Power Cord IEC provides a power cord with each Micromax centrifuge.
The unit requires a grounded power supply (3-prong power
outlet). If your facility does not have properly grounded
power outlets, arrange for proper grounding. Do not
remove the grounding pin from the centrifuge power cord.
Do not use an adapter to connect to a 2-prong outlet.
Micromax Series
4
Service Manual