Thermo Scientific APEX Quick-Start Guide
REC-F-58E—Part #085555 Page 1
APEX Metal
Detector Quick-
Start Guide
Control Panel
The main components of the detector’s
control panel are as follows.
The display panel.
The three (red, green, and yellow)
indicator lights.
The green Go button.
The red Back button.
The four blue triangular-shaped
navigation buttons.
The Indicator Lights
These give you a quick overview of how the
detector is functioning.
Red—Flashing once indicates excess
product effect; steady indicates a fault.
Green—Indicates a product is present
in the detector’s search head.
Yellow—Indicates a contaminant has
been detected in the product.
The Navigation Buttons
These allow you to navigate around the
detector’s menus and sub-menus and are
used to increase or decrease numbers and
select characters and settings in various
menus and input screens.
The Go Button
Press the Go button to select or start one of
the detector’s functions.
The Back Button
Press the Back button to stop one of the
detector’s functions or to return to the
previous menu screen. Press repeatedly to
return to the Main Menu.
Main Menu
The APEX’s Main Menu normally looks like
this.
Product Name
However, if you have enabled the IXR
function (see page 8), the Main Menu looks
like this.
Product Name
Naming the Product
This allows you to name the product you are
testing.
1) Make sure the Main Menu is displayed.
2) Press the Go button and the product-
calibration menu appears.
3) Press the down-navigation button to
highlight the “Edit Product Name”
function.
Edit Product Name

Other manuals for Thermo Scientific Apex

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Apex and is the answer not in the manual?

Thermo Scientific Apex Specifications

General IconGeneral
Power Requirements100-240 VAC, 50/60 Hz
ConstructionStainless steel
Sensitivity Non-FerrousUp to 1.0 mm
IP RatingIP65
Line Speed60 m/min
Operating Temperature0°C to 40°C
Conveyor Speed60 m/min
OutputsRelay

Summary

Control Panel Overview

Indicator Lights

Red, green, yellow lights indicate detector status and product detection.

Go Button

Used to select or start detector functions.

Conveyor Application Setup

Reject Device Configuration

Noise Threshold Calibration

Establishing Detector Noise Baseline

Calibrates detector noise baseline with conveyor running, no product.

Full Product Calibration

Learning Uncontaminated Product

Detector learns product, setting phase-angle and detection level.

Pipeline Application Setup

Rx Application Setup

IntelliTrack XR™ (IXR) Functionality

IXR Detects Stainless Steel

Detects stainless steel and metallic contaminants in wet/dry products.

IXR Prerequisites and Usage

Requires photo-eye and specific parameters for IXR to function.

Enabling IntelliTrack XR™ (IXR)

Activating the IXR Feature

Steps to enable the IXR function via the system and tools menu.

Calibrating the IXR Function

IXR Calibration Process

Calibrates noise thresholds and detect levels for IXR.

Auto-Calibration of IXR Function

Optimizing IXR Auto-Calibration Process

Procedures for optimizing auto-calibration with specific conditions.