Cimarec
Hot Plates
OPERATION MANUAL
AND PARTS LIST
Model # Top Plate Size Voltage
HP130914 4x4 100
HP130915 4x4 120
HP130910-33 4x4 220-240
HP131224 7x7 100
HP131225 7x7 120
HP131220-33 7x7 220-240
HP131534 10x10 100
HP131535 10x10 120
HP131530-33 10x10 220-240
1
LT1312X1 10/28/09

Questions and Answers

  • A
    April GutierrezAug 7, 2025
    What to do if my Thermo Scientific Cimarec HP131535 shows error message E02?
    • E
      Ellen RobinsonAug 7, 2025
      If your Thermo Scientific Cooktop displays the message 'E02 Call Tech. Service', it could be due to a few reasons. It might be that the thermocouple isn't connected, in which case you should replace the element assembly attached to the thermocouple. Another possibility is a thermocouple open circuit, requiring you to replace the element. Finally, the control board might need replacement.
  • J
    Jose CantuAug 21, 2025
    What does error E01 mean on my Thermo Scientific Cooktop and how to fix it?
    • D
      Dylan MossAug 21, 2025
      If your Thermo Scientific Cooktop displays the message 'E01 Call Tech. Service', it could be due to a thermocouple that is not connected, in which case you should replace the element assembly attached to the thermocouple. It could also be caused by a thermocouple open circuit, requiring you to remove the short.

Summary

Safety Information

Alert Signals

Icons indicating warning, caution, and notes for user safety and equipment protection.

Warnings

Crucial safety warnings related to electrical shock, personal injury, and hazardous materials.

Specific Safety Precautions

Detailed safety advice on cleaning, handling, cord integrity, and operating environment.

Specifications

Introduction

Principles of Operation

Unpacking and Installation

Operation

Maintenance

Troubleshooting

Error Codes

Identification and potential causes/solutions for displayed error codes (E01, E02).

Replacement Parts

Thermo Scientific Cimarec HP131535 Specifications

General IconGeneral
ModelHP131535
TypeHotplate
Control TypeAnalog
Voltage120 V
Frequency60 Hz
Maximum Temperature540°C
Temperature RangeUp to 540°C
Electrical Requirements120V 60Hz