EasyManua.ls Logo

Thermo Scientific SUPER-NUOVA+ SP88857190 - User Manual

Thermo Scientific SUPER-NUOVA+ SP88857190
43 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
1
SUPER-NUOVA
+
Hot Plates, Stirrers and
Stirring Hotplates
OPERATION MANUAL
AND PARTS LIST
Thermo
SCIENTIFIC
CIC0000810 • 03/09/15

Questions and Answers

  • Z
    zprestonAug 9, 2025
    What to do if my Thermo Scientific SUPER-NUOVA+ SP88857190 Cooktop displays 'Check Heater'?
    • R
      robert43Aug 9, 2025
      If your Thermo Scientific Cooktop displays 'Check Heater', it could be due to several reasons. It might be an internal thermocouple short circuit, a failure in the internal thermocouple or the element, or a failure in the optocoupler/triac circuit. To resolve this, try the following: remove the short, replace the thermocouple attached to the element, replace the element, or replace the control board.
  • R
    Rebecca CraigAug 15, 2025
    Why does my Thermo Scientific SUPER-NUOVA+ SP88857190 Cooktop show 'Check Motor'?
    • M
      Mary TurnerAug 15, 2025
      If your Thermo Scientific Cooktop displays 'Check Motor', it could be due to a locked rotor condition, failure of the motor, failure of the motor circuit, or failure of the motor sensor. Here are some steps to troubleshoot the issue: * Free the locked rotor. * Replace the motor. * Replace the control board. * Replace the motor sensor.
  • L
    leoncindyAug 20, 2025
    What to do if Thermo Scientific SUPER-NUOVA+ SP88857190 Cooktop shows 'Over Temp'?
    • L
      latoyanewmanAug 20, 2025
      If your Thermo Scientific Cooktop displays 'Over Temp', it is likely due to an OTP circuit failure or the OTP potentiometer being set too low. The suggested solutions are to replace the control board or increase the OTP setting.
  • R
    Rebecca ThomasAug 27, 2025
    What to do if my Thermo Scientific SUPER-NUOVA+ SP88857190 Cooktop displays 'InProbe Err'?
    • C
      cbutlerAug 27, 2025
      If your Thermo Scientific Cooktop displays 'InProbe Err', it indicates a problem with the internal thermocouple. This could be due to the thermocouple not being connected, the thermocouple being open, or the thermocouple being connected backwards (reversed polarity). To fix this, ensure proper connection and polarity of the thermocouple, replace the thermocouple attached to the element, and ensure proper connection and polarity.
  • B
    Billy WellsSep 1, 2025
    How to fix 'ExProbe Err' on Thermo Scientific SUPER-NUOVA+ SP88857190 Cooktop?
    • V
      Valerie MartinSep 1, 2025
      If your Thermo Scientific Cooktop displays 'ExProbe Err', it means the external probe is either left out of the solution or connected backwards. To resolve this, place the external probe into the solution or correct the orientation of the external probe.

Summary

Safety Information

Alert Signals and Warnings

Explains safety alert symbols and provides general warnings for safe operation.

General Specifications

Unpacking and Installation

Unpacking Procedures

Outlines the steps for safely unpacking the unit and checking included items.

Installation Guidelines

Provides instructions for properly setting up and connecting the unit to a power source.

Principles of Operation

Operation

Setting the Stirring Speed

Details the procedure for setting and adjusting the stirring speed using the control panel.

Setting the Temperature

Guides users on how to set and control the temperature using the control panel and probes.

Controlling Solution Temperature with External Probe

Explains how to use an external probe for precise solution temperature control.

Using Preset Programs

Describes how to program and activate preset temperature and speed settings.

Hardware Over-Temperature Protection

Details the hardware OTP feature for preventing overheating and its adjustment.

System Configuration

Troubleshooting

Error Codes and Solutions

Lists common error messages, their causes, and recommended solutions for the unit.

Replacement Parts List

Thermo Scientific SUPER-NUOVA+ SP88857190 Specifications

General IconGeneral
BrandThermo Scientific
ModelSUPER-NUOVA+ SP88857190
CategoryCooktop
LanguageEnglish

Related product manuals