What to do if alarm LED stays on and control is higher than set temperature in Thermo Scientific Oven?
- TTyler MyersAug 6, 2025
If the alarm LED remains illuminated and the control indicates a temperature exceeding the set point, it could be due to several reasons. The set temperature might have been adjusted to a value lower than the actual temperature, minus the high alarm limit, in which case you should wait for the actual temperature to cool down to the set temperature. Alternatively, there may be a defective controller, which would require replacement. Finally, defective solid state relays or safety relays could also be the cause; in this case, refer to the schematic and replace the relays.